Practice and applications of 17-O-excess measurements of water using novel laser spectroscopy
Abstract:
In natural waters, variability in 17O-excess is very small (on the order of tens of per meg, where 1 per meg is 0.001‰). Until recently, only measurements made via Isotope Ratio Mass Spectrometry (IRMS) could achieve the required precision, following time-consuming front-end chemistry that converted H2O into O2 for analysis of m/z+ 32, 33 and 34. Recent improvements in laser-based spectroscopy, e.g., Cavity Ring-Down Spectroscopy (CRDS), are enabling quicker and easier measurement of 17O-excess in water (Steig et al., 2013; 2014). The Picarro L2140-i is certified with a precision of ≤ 0.015‰; however implementation of best practices can result in an achievable precision of ≤ 0.008‰, thereby demonstrating comparable performance to IRMS. We will review our recommendations for achieving high-precision measurements of 17O-excess on the Picarro L2140-i, including how to calibrate the system, the frequency of standards analysis, the number of replicate injections and vials required, and approaches to dealing with sample-to-sample memory. We will also compare the external accuracy achieved by three distinct Picarro L2140-i analyzers for multiple waters with distinct isotopic composition.