The electrolytic-cell-based polarographic technique involves placing a contact lens on a curved electrode assembly, which has a gold cathode (working electrode) at the center and silver anode surrounding it. Voltage-current data are used to fix limiting voltage across the electrodes so as to maintain limiting current (mass transfer regime) conditions. A steady state experiment is carried out at physiological temperature (35oC) to determine the oxygen permeability (Dk) through the contact lenses. The apparatus is calibrated with a contact lens of known oxygen permeability. An unsteady-state experiment is performed to determine both diffusivity (D) and solubility (k) in the same experiment. Here, the calibration is done using a membrane of known oxygen diffusivity and solubility. For the first time, values of D and k are determined for new high-Dk silicone-hydrogel soft contact lenses as well as more traditional HEMA-based soft contact lenses.