506878 Point-of-Care Detection of Hematocrit in a Microfluidic System Integrating Microfabricated Carbon Electrodes

Hwi Yong Lee, Jessika A. Rogers, Chito Kendrick and Adrienne R. Minerick, Chemical Engineering, Michigan Technological University, Houghton, MI

This work describes an electrochemical microfluidic system for point-of-care detection of hematocrit. Hematocrit, the volume percentage of red blood cells in whole blood, is an important medical indicator. Its deviations from normal ranges (40–54% for men and 36–48% for women) are symptoms of various medical conditions such as anemia and leukemia.

We fabricated carbon microelectrodes via a microfabrication strategy. Carbon precursors, negative photoresist SU-8, were photolithographically patterned on quartz substrates and pyrolyzed at 900 OC. Specific pyrolysis conditions were optimized to obtain carbon electrodes from large SU-8 features (greater than 10 mm in length) containing more internal stress and residual gases. The SU-8 features were hard-baked at 200 OC under vacuum pumping to completely evaporate residual gases and to enhance the adhesion. Then, the hard-baked SU-8 were pyrolyzed via a conventional two-step pyrolysis: 1) hard-baking at 200 OC for 30 mins and 2) pyrolysis at 900 OC for 1 hour. SU-8 pyrolyzed carbon electrodes were combined with a single PDMS microchannel to complete a device.

To test the feasibility for detection of hematocrit, current responses of red blood cells suspended in phosphate buffered saline were measured at 100 V for 30 secs. A linear trend line of current versus red blood cell concentration was obtained. The results indicate that our device can determine hematocrit within 3.67 % precision and 3.97 % accuracy. Excellent performance of carbon electrodes was attributed to their surface properties. Compared to metal electrodes, carbon electrodes have less bio-fouling tendencies. Large electrode surface areas with porous structures also increase the system voltage efficiency by decreasing the charge transfer overpotential at electrode/electrolyte interfaces.


Extended Abstract: File Not Uploaded