Modeling and Testing of a New Sternal Closure Device Using Tubular Mesh-like Superelastic Nitinol Structure
Mr. Yannick-Vincent Baril1, Prof. Vladimir Brailovski1, Prof. Patrick Terriault1 and Dr. Raymond Cartier2, 1Mechanical Engineering, Ecole de technologie superieure, 1100, Notre-Dame Street West, Montreal, QC H3C 1K3, Canada, 2Interventional Cardiology, Montreal Heart Institute, 5000, Belanger Street, Montreal, QC H1T 1C8, Canada

To reduce a risk of sternal dehiscence of the patients subjected to median sternotomy, a new sternal closure device is developed. This device prevents the cut in and through the sternum bones during aggressive post-surgery physiological activities as coughing, deep breathing and sudden movement. Furthermore, the new device is capable of maintaining a nearly constant pressure between sternum halves during rehabilitation of such patients, thus improving the bone healing conditions. The principal element of such a device consists in a tubular mesh-like structure made of superelastic Nitinol, which tends towards a flat form when it is in contact with the sternum. Finite element modeling and in-vitro laboratory testing of the tubular superelastic device confirm a 30% reduction in contact pressure exerted by this device on the sternum bones, and a 25% increase in compressive pressure between the sternum halves, as compared to the conventional sternal steel wire closure device.

Session 10: Device Modeling and Testing