Impact of Testing Methodology on Breakdown Potentials
Dr. Clifford P. Warner1 and Mr. Richard A. Corbett2, 1W. L. Gore & Associates, P.O. Box 1320, Elkton, MD 21922-1320, 2Corrosion Testing Laboratories, Inc., 60 Blue Hen Drive, Newark, DE 19713

The ASTM F2129 test method is a valuable method for determining the susceptibility of small medical implants to localized corrosion.  However, one limitation of its current form, particularly at a scan rate of 0.1667mV/sec, is testing time per specimen.  Several modifications to the ASTM F2129 method were evaluated for their impact on breakdown potentials, including: scan rates, open circuit hold method and elimination of the 100mV cathodic excursion at the start of the test.  Experiments were performed on commercially available stainless steel and nitinol wires.