The sampling concept was demonstrated using a test vessel whose geometry mimicked a 5 cc stirred micro reactor. Two modes of liquid sampling were evaluated: (1) flashing of the liquid sample into a calibrated loop of a 6-port ValcoŽ gas sampling valve; and (2) flashing of the liquid sample directly into the injection port of a HP 5890 Series II gas chromatograph. The test samples consisted of an aqueous mixture of ca. 1 % each of C1 to C4 alcohols, and a crude reaction product obtained from a commercial-scale cyclohexane oxidation reactor.
The results showed that the ability to fill the sampling loop following conventional practice was dependent upon the pressure developed by the flashing liquid. This effect was not observed for the direct injection method developed here since the total GC area counts were reproducible within 0.1 %. Details on the valve design, test results, and its application to the oxidation of para-xylene to terephthalic acid will be described.